One of the most difficult problems facing the instrument engineer is the accurate calibration of orifice meters, particularly at remote or inaccessible locations. The object of this paper is to describe a unique solution to this problem, an automatic pneumatic deadweight test utilizing the floating ball principle. The Ametek Model PK Pneumatic Deadweight Tester is an ideal instrument to use for the calibration of orifice meters. This instrument provides unexcelled accuracy combined with reliable operation and ease of use for both field and laboratory calibration. It has been accepted by most of the gas transmission companies as a pressure standard for meter calibration.


    Access

    Access via TIB

    Check availability in my library


    Export, share and cite



    Title :

    Instrument calibration using the pneumatic deadweight tester


    Additional title:

    Eichung von Messinstrumenten mit pneumatischem Eigengewichtspruefgeraet


    Contributors:


    Publication date :

    1983


    Size :

    3 Seiten, 3 Bilder


    Type of media :

    Conference paper


    Type of material :

    Print


    Language :

    English




    Miniature Remote Deadweight Calibrator