Vibrating columns of fluid used to calibrate transducers. Dynamic pressure calibration standard developed for calibrating flush diaphragm-mounted pressure transducers. Pressures up to 20 kPa (3 psi) accurately generated over frequency range of 50 to 1,800 Hz. System includes two conically shaped aluminum columns one 5 cm (2 in.) high for low pressures and another 11 cm (4.3 in.) high for higher pressures, each filled with viscous fluid. Each column mounted on armature of vibration exciter, which imparts sinusoidally varying acceleration to fluid column. Signal noise low, and waveform highly dependent on quality of drive signal in vibration exciter.


    Access

    Access via TIB

    Check availability in my library


    Export, share and cite



    Title :

    Dynamic Pressure Calibration Standard


    Contributors:
    Schutte, P. C. (author) / Cate, K. H. (author) / Young, S. D. (author)

    Published in:

    Publication date :

    1986-06-01



    Type of media :

    Miscellaneous


    Type of material :

    No indication


    Language :

    English


    Keywords :


    Dynamic Calibration of Pressure Transducers