Described is a potentiometric system for measurement of a 25-ohm platinum capsule-type standard over the range 20-100 k, a germanium standard over the range 4-20 k, and 1380-ohm platinum test sensors over the range 4-100 k. By means of function and output switches thermometer identification, thermometer resistances, and static data are read out automatically on a typewriter. Control of temperature to 0.01 k and optimum uniformity between thermometers is required. Other requirements are overall calibration accuracy of no worse than &0.2 k, with design targets of &0.005 k uncertainty from temperature as determined by the standards.


    Zugriff

    Zugriff über TIB

    Verfügbarkeit in meiner Bibliothek prüfen


    Exportieren, teilen und zitieren



    Titel :

    Cryogenic temperature sensor calibration with automated data readout


    Beteiligte:
    Daneman, H.L. (Autor:in)

    Erschienen in:

    ISA Transactions ; 8 , 2 ; 151-156


    Erscheinungsdatum :

    1969


    Format / Umfang :

    6 Seiten, 7 Bilder, 2 Tabellen


    Medientyp :

    Aufsatz (Zeitschrift)


    Format :

    Print


    Sprache :

    Englisch