Electrical measurements made by four point probes, two of which double as temperature probes. Laboratory apparatus measures both Seebeck coefficients and electrical resistivities of candidate thermoelectric materials at temperatures from ambient to 1,300 K. Apparatus makes possible to take both measurements alternately and in rapid succession during same heating cycle, thereby reducing distortion.
Apparatus Measures Seebeck Coefficient And Resistivity
NASA Tech Briefs ; 17 , 5
1993-05-01