This paper reports a laser diode pumped power build-up cavity (PBC) to apply for cavity ringdown spectroscopy (CRDS). The PBC is a compact-high finesse external cavity laser diode, which can obtain resonant enhancement without active cavity length control. PBC is applied generally undesired feed backed beam to cavity resonant enhancement. More than 2.30 /spl mu/s of ringdown time was achieved with a very small gas sampling volume about 2.5ml. Sensitivity of 3.62 /spl times/ 10/sup -7/ cm/sup -1/ corresponding to NO/sub 2/ detection limit of less than 1.7 ppm was achieved.
Novel NO/sub 2/ detection based on cavity ringdown spectroscopy using a power build-up cavity
2005-01-01
172325 byte
Conference paper
Electronic Resource
English
Novel NO~2 detection based on Cavity Ringdown Spectroscopy using a Power Build-up Cavity
British Library Conference Proceedings | 2005