Purpose We present the validation of PROST, a robotic device for prostate biopsy. PROST is designed to minimize human error by introducing some autonomy in the execution of the key steps of the procedure, i.e., target selection, image fusion and needle positioning. The robot allows executing a targeted biopsy through ultrasound (US) guidance and fusion with magnetic resonance (MR) images, where the target was defined. Methods PROST is a parallel robot with 4 degrees of freedom (DOF) to orient the needle and 1 DOF to rotate the US probe. We reached a calibration error of less than 2 mm, computed as the difference between the needle positioning in robot coordinates and in the US image. The autonomy of the robot is given by the image analysis software, which employs deep learning techniques, the integrated image fusion algorithms and automatic computation of the needle trajectory. For safety reasons, the insertion of the needle is assigned to the doctor. Results System performance was evaluated in terms of positioning accuracy. Tests were performed on a 3D printed object with nine 2-mm spherical targets and on an anatomical commercial phantom that simulates human prostate with three lesions and the surrounding structures. The average accuracy reached in the laboratory experiments was 1.30 ± 0.44 mm in the first test and 1.54 ± 0.34 mm in the second test. Conclusions We introduced a first prototype of a prostate biopsy robot that has the potential to increase the detection of clinically significant prostate cancer and, by including some level of autonomy, to simplify the procedure, to reduce human errors and shorten training time. The use of a robot for the biopsy of the prostate will create the possibility to include also a treatment, such as focal ablation, to be delivered through the same system.


    Access

    Download


    Export, share and cite



    Title :

    Toward autonomous robotic prostate biopsy: a pilot study



    Publication date :

    2021-01-01



    Type of media :

    Article (Journal)


    Type of material :

    Electronic Resource


    Language :

    English



    Classification :

    DDC:    629



    Autonomous Robotic System for Breast Biopsy With Deformation Compensation

    Ferrari, Sandro / Tagliabue, Eleonora / Maris, Bogdan Mihai et al. | BASE | 2023

    Free access

    A Virtual Reality System for Pre-Planning of Robotic-Assisted Prostate Biopsy

    Gîrbacia, Florin ;Butnariu, Silviu ;Voinea, Daniel | Trans Tech Publications | 2015


    Preclinical Validation of a Semi-Autonomous Robot for Transperineal Prostate Biopsy

    Bogdan Maris / Maria-Camilla Fiazza / Michela De Piccoli et al. | BASE | 2022

    Free access

    Robotic prostate surgery

    Muntener, Michael / Ursu, Daniel / Patriciu, Alexandru et al. | Taylor & Francis Verlag | 2006


    Planning of Needle Insertion for Robotic-Assisted Prostate Biopsy in Augmented Reality Using RGB-D Camera

    Gîrbacia, Florin / Boboc, Răzvan / Gherman, Bogdan et al. | Springer Verlag | 2016