In photometry and radiometry, photodetectors such as silicon detector and PMT detector are widely used. In precision metrology, the uncertainty of the nonlinearity should be considered. Superposition method is used to analyse the linearity. The silicon trap detector is measured using both nonmochromator light and monochromator light. First, integrating sphere with broadband light is used to test the linearity. The result shows that the nonlinearity is (1-3)×10-4 from 1uA to 1mA. The monchromator light result shows that the nonlinearity is below 3×10-4 from 1uA to 1mA, which is consistent with the integrating sphere method. For the PMT detector, the linearity is measured only using monochromator light. Experiment shows that the nonlinearity is less than 1×10-3 through three orders of magnitude.
Nonlinearity measurement of silicon detector and PMT detector
AOPC 2017: Space Optics and Earth Imaging and Space Navigation ; 2017 ; Beijing,China
Proc. SPIE ; 10463
2017-10-24
Aufsatz (Konferenz)
Elektronische Ressource
Englisch
Nonlinearity measurement of silicon detector and PMT detector [10463-76]
British Library Conference Proceedings | 2017
Multiplexed extrinsic silicon detector array
NTRS | 1982
|Interferometric Hetero-Detector Phase Measurement
British Library Online Contents | 2004
|Silicon-detector arrays advance medical imaging
British Library Online Contents | 1993
|Range-measurement-type optical vehicle detector
Kraftfahrwesen | 1995
|