Basic investigations, consisting primarily of airflow measurements and visual smoke tests, were made on four commercially available units. The results indicated that performance varied considerably between units. Therefore, it was decided to outline the performance desired and to include in the overall equipment specifications a set of performance tests to assure a clean air safety cabinet with respect to chemical usage. The evaluation was to be based on aerodynamic performance under the various modes of operation.