The design of a gyrotron for continuous wave operation at 395 GHz at a power level from several watts to 2 kilowatts is presented. The gyrotron oscillator will be used for development of high sensitive NMR with dynamic nuclear polarization (DNP) method. The gyrotron design is optimized for single mode operation at second harmonic of the electron cyclotron frequency, ¿=2¿c by removing the mode competition with fundamental operations (¿=2¿c).


    Zugriff

    Zugriff prüfen

    Verfügbarkeit in meiner Bibliothek prüfen

    Bestellung bei Subito €


    Exportieren, teilen und zitieren



    Titel :

    Design of a 400 GHz Gyrotron for DNP-NMR Spectroscopy


    Beteiligte:
    La Agusu, (Autor:in) / Murase, H. (Autor:in) / Idehara, T. (Autor:in) / Saito, T. (Autor:in) / Mitsudo, S. (Autor:in) / Takahashi, D. (Autor:in) / Fujiwara, T. (Autor:in)


    Erscheinungsdatum :

    2006-09-01


    Format / Umfang :

    928443 byte




    Medientyp :

    Aufsatz (Konferenz)


    Format :

    Elektronische Ressource


    Sprache :

    Englisch



    Conceptual Design Study of a Novel Gyrotron for NMR/DNP Spectroscopy

    Sabchevski, S. / Idehara, T. / Mitsudo, S. et al. | British Library Online Contents | 2005


    The Design of the 394.6 Ghz Continuously Tunable Coaxial Gyrotron for DNP Spectroscopy

    Glyavin, M. / Khizhnyak, V. / Luchinin, A. et al. | British Library Online Contents | 2008


    Design of the ITER 170 GHz gyrotron gun

    Yovchev, I. / Hogge, J.-P. / Goodman, T. et al. | IEEE | 2004


    Design of a CW 1 THz Gyrotron (Gyrotron Fu Cw III) Using a 20 T Superconducting Magnet

    Agusu, L. / Idehara, T. / Mori, H. et al. | British Library Online Contents | 2007


    8mm TE13 Mode Gyrotron