Using the in-flight calibration information from the Coastal Zone Colour Scanner(CZCS) housekeeping data for various dates, it is shown that the radiance from the calibration lamp is changing considerably with time. This changing radiance is shown to be consistent with a change in the temperature of the calibration lamp from its nominal value of 2000 K. Thus there is a need to find a way to determine the correct value of the radiance from the calibration lamp.
Coastal zone colour scanner: faiilure of active calibration
Fehler der selbsttaetigen Eichung bei der Meerwasserfarbmessung von Satelliten
Journal of Physics E. Scientific Instruments ; 15 , 10 ; 1003-1007
1982
5 Seiten, 3 Bilder, 3 Tabellen, 7 Quellen
Aufsatz (Zeitschrift)
Englisch
Algorithms for Coastal-Zone Color-Scanner Data
NTRS | 1986
Ocean color imagery: Coastal zone color scanner
NTRS | 1975
|Scanner calibration method, device and system and crane
Europäisches Patentamt | 2023
|