Study Design: Biomechanical study. Objective: Recently, a posterior concave periapical distraction device for fusionless scoliosis correction was introduced. The goal of this study was to quantify the effect of the periapical distraction device on spinal range of motion (ROM) in comparison with traditional rigid pedicle screw-rod instrumentation. Methods: Using a spinal motion simulator, 6 human spines were loaded with 4 N m and 6 porcine spines with 2 N m to induce flexion-extension (FE), lateral bending (LB), and axial rotation (AR). ROM was measured in 3 conditions: untreated, periapical distraction device, and rigid pedicle screw-rod instrumentation. Results: The periapical distraction device caused a significant (P <.05) decrease in ROM of FE (human, −40.0% and porcine, −55.9%) and LB (human, −18.2% and porcine, −17.9%) as compared to the untreated spine, while ROM of AR remained unaffected. In comparison, rigid instrumentation caused a significantly (P <.05) larger decrease in ROM of FE (human, −80.9% and porcine, −94.0%), LB (human, −75.0% and porcine, −92.2%), and AR (human, −71.3% and porcine, −86.9%). Conclusions: Although no destructive forces were applied, no device failures were observed. Spinal ROM was significantly less constrained by the periapical distraction device compared to rigid pedicle screw-rod instrumentation. Therefore, provided that scoliosis correction is achieved, a more physiological spinal motion is expected after scoliosis correction with the posterior concave periapical distraction device.


    Zugriff

    Download


    Exportieren, teilen und zitieren



    Titel :

    A Novel Spinal Implant for Fusionless Scoliosis Correction:A Biomechanical Analysis of the Motion Preserving Properties of a Posterior Periapical Concave Distraction Device


    Beteiligte:

    Erscheinungsdatum :

    2017-08-01


    Anmerkungen:

    Holewijn , R M , de Kleuver , M , van der Veen , A J , Emanuel , K S , Bisschop , A , Stadhouder , A , van Royen , B J & Kingma , I 2017 , ' A Novel Spinal Implant for Fusionless Scoliosis Correction : A Biomechanical Analysis of the Motion Preserving Properties of a Posterior Periapical Concave Distraction Device ' , Global Spine Journal , vol. 7 , no. 5 , pp. 400-409 . https://doi.org/10.1177/2192568217699377



    Medientyp :

    Aufsatz (Zeitschrift)


    Format :

    Elektronische Ressource


    Sprache :

    Englisch



    Klassifikation :

    DDC:    629



    The biomechanical fundamentals of crosslink-augmentation in posterior spinal instrumentation

    Cornaz, Frédéric / Fasser, Marie-Rosa / Snedeker, Jess Gerrit et al. | BASE | 2022

    Freier Zugriff

    DISTRACTION ASSESSMENT DEVICE AND DISTRACTION ASSESSMENT METHOD

    SAKAI YASUTOSHI / HAYASHI KEITA / TANIYAMA HIROSHI | Europäisches Patentamt | 2022

    Freier Zugriff

    DISTRACTION ASSESSMENT DEVICE AND DISTRACTION ASSESSMENT METHOD

    SAKAI YASUTOSHI / HAYASHI KEITA / TANIYAMA HIROSHI | Europäisches Patentamt | 2022

    Freier Zugriff

    Cervical Spinal Cord Injury Using Biomechanical Experimentation

    Yoganandan, N. / Pintar, F. A. / Maiman, D. J. et al. | British Library Conference Proceedings | 1995


    Cervical spinal cord injury using biomechanical experimentation

    Yoganandan,N. / Pintar,F.A. / Maiman,D.J. et al. | Kraftfahrwesen | 1995