In phase-resolved OCT angiography blood flow is detected from phase changes in between A-scans that are obtained from the same location. In ophthalmology, this technique is vulnerable to eye motion. We address this problem by combining inter-B-scan phase-resolved OCT angiography with real-time eye tracking. A tracking scanning laser ophthalmoscope (TSLO) at 840 nm provided eye tracking functionality and was combined with a phase-stabilized optical frequency domain imaging (OFDI) system at 1040 nm. Real-time eye tracking corrected eye drift and prevented discontinuity artifacts from (micro)saccadic eye motion in OCT angiograms. This improved the OCT spot stability on the retina and consequently reduced the phase-noise, thereby enabling the detection of slower blood flows by extending the inter-B-scan time interval. In addition, eye tracking enabled the easy compounding of multiple data sets from the fovea of a healthy volunteer to create high-quality eye motion artifact-free angiograms. High-quality images are presented of two distinct layers of vasculature in the retina and the dense vasculature of the choroid. Additionally we present, for the first time, a phase-resolved OCT angiogram of the mesh-like network of the choriocapillaris containing typical pore openings. © 2012 Optical Society of America.


    Zugriff

    Download


    Exportieren, teilen und zitieren



    Titel :

    Real-time eye motion correction in phase-resolved OCT angiography with tracking SLO


    Beteiligte:
    Braaf, B. (Autor:in) / Vienola, K.V. (Autor:in) / Sheehy, C.K. (Autor:in) / Yang, Q. (Autor:in) / Vermeer, K.A. (Autor:in) / Tiruveedhula, P. (Autor:in) / Arathorn, D.W. (Autor:in) / Roorda, A.J. (Autor:in) / de Boer, J.F. (Autor:in)

    Erscheinungsdatum :

    2013-01-01


    Anmerkungen:

    Braaf , B , Vienola , K V , Sheehy , C K , Yang , Q , Vermeer , K A , Tiruveedhula , P , Arathorn , D W , Roorda , A J & de Boer , J F 2013 , ' Real-time eye motion correction in phase-resolved OCT angiography with tracking SLO ' , Biomedical Optics Express , vol. 4 , pp. 51-65 . https://doi.org/10.1364/BOE.4.000051



    Medientyp :

    Aufsatz (Zeitschrift)


    Format :

    Elektronische Ressource


    Sprache :

    Englisch


    Klassifikation :

    DDC:    629



    Optimized respiratory-resolved motion-compensated 3D Cartesian coronary MR angiography

    Correia, Teresa / Ginami, Giulia / Cruz, Gastão et al. | BASE | 2018

    Freier Zugriff

    Tracking Vision: A real-time motion tracking system

    British Library Online Contents | 1998


    Optimal Polyline Tracking for Artery Motion Compensation in Coronary Angiography

    Dubuisson-Jolly, M.-P. / Liang, C.-C. / Gupta, A. et al. | British Library Conference Proceedings | 1998


    Correction of motion tracking errors for PET head rigid motion correction

    Miranda Menchaca, Alan / Kroll, Tina / Schweda, Vanessa et al. | BASE | 2023

    Freier Zugriff

    Motion compensated digital subtraction angiography

    Hemmendorff, M. / Knutsson, Hans / Andersson, Mats et al. | BASE | 1999

    Freier Zugriff