OBJECTIVES: To evaluate the impact of movement and motion-artefact correction systems on CBCT image quality and interpretability of simulated diagnostic tasks for aligned and lateral-offset detectors. METHODS: A human skull simulating three diagnostic tasks (implant planning in the anterior maxilla, implant planning in the left-side-mandible and mandibular molar furcation assessment in the right-side-mandible) was mounted on a robot performing six movement types. Four CBCT units were used: Cranex 3Dx (CRA), Ortophos SL (ORT), Promax 3D Mid (PRO), and X1. Protocols were tested with aligned (CRA, ORT, PRO, and X1) and lateral-offset (CRA and PRO) detectors and two motion-artefact correction systems (PRO and X1). Movements were performed at one moment-in-time (t1), for units with an aligned detector, and three moments-in-time (t1-first-half of the acquisition, t2-second-half, t3-both) for the units with a lateral-offset detector. 98 volumes were acquired. Images were scored by three observers, blinded to the unit and presence of movement, for motion-related stripe artefacts, overall unsharpness, and interpretability. Fleiss' κ was used to assess interobserver agreement. RESULTS: Interobserver agreement was substantial for all parameters (0.66-0.68). For aligned detectors, in all diagnostic tasks a motion-artefact correction system influenced image interpretability. For lateral-offset detectors, the interpretability varied according to the unit and moment-in-time, in which the movement was performed. PRO motion-artefact correction system was less effective for the offset detector than its aligned counterpart. CONCLUSION: Motion-artefact correction systems enhanced image quality and interpretability for units with aligned detectors but were less effective for those with lateral-offset detectors.


    Zugriff

    Download


    Exportieren, teilen und zitieren



    Titel :

    Impact of movement and motion-artefact correction on image quality and interpretability in CBCT units with aligned and lateral-offset detectors


    Beteiligte:

    Erscheinungsdatum :

    2020-01-01


    Anmerkungen:

    Santaella , G M , Wenzel , A , Haiter-Neto , F , Rosalen , P L & Spin-Neto , R 2020 , ' Impact of movement and motion-artefact correction on image quality and interpretability in CBCT units with aligned and lateral-offset detectors ' , Dento maxillo facial radiology , vol. 49 , no. 1 , 20190240 . https://doi.org/10.1259/dmfr.20190240



    Medientyp :

    Aufsatz (Zeitschrift)


    Format :

    Elektronische Ressource


    Sprache :

    Englisch



    Klassifikation :

    DDC:    629



    An ex vivo study of automated motion artefact correction and the impact on cone beam CT image quality and interpretability

    Spin-Neto, Rubens / Matzen, Louise H / Schropp, Lars W et al. | BASE | 2018

    Freier Zugriff

    Motion control for CBCT gantry

    Karppi, Harri | BASE | 2019

    Freier Zugriff

    Respiratory motion model derived from CBCT projection data

    Akintonde, A / Thielemans, K / Sharma, R et al. | BASE | 2019

    Freier Zugriff

    Gait Reconstruction from Motion Artefact Corrupted Fabric-Embedded Sensors

    Michael, Brendan / Howard, Matthew Jacob William | BASE | 2018

    Freier Zugriff

    ARTEFACT WITH HEATER FILM

    GUADAGNO LIBERATA / VERTUCCIO LUIGI / LONGO RAFFAELE et al. | Europäisches Patentamt | 2024

    Freier Zugriff